Skip Navigation
Skip to contents

ACC : Acute and Critical Care

OPEN ACCESS
SEARCH
Search

Search

Page Path
HOME > Search
1 "equipment design"
Filter
Filter
Article category
Keywords
Publication year
Authors
Original Article
Pulmonary
The Effect of Positive End-Expiratory Pressure on Air Leakage: Comparison of Cuff Designs
Junyong In, Gyung Serk Shim, Seunghyun Chung
Korean J Crit Care Med. 2014;29(1):3-6.   Published online February 28, 2014
DOI: https://doi.org/10.4266/kjccm.2014.29.1.3
  • 4,865 View
  • 161 Download
  • 2 Crossref
AbstractAbstract PDF
Background
Recently developed taper-shaped cuffs (TG cuffs) of endotracheal tubes (ETTs) are known to have a more potent sealing effect than cylindrical high-volume low-pressure cuffs (HL cuffs) of conventional ETTs. The aim of this study was to compare TG cuffs with HL cuffs of ETTs in a bench-top model with regard to air leakage under various positive end-expiratory pressures (PEEP).
Methods
HL cuffs and TG cuffs made from PVC were included (HL group vs. TG group). A model trachea with an internal diameter (ID) of 22 mm was attached to a test lung. The test lung was ventilated using an anesthesia respirator with volume controlled mode and PEEPs of 0, 5, 10, or 15 cm H2O. Using spirometry, percentages of expired to inspired tidal volumes (TVe/i) were calculated as a measure of air leakage.
Results
With regard to PEEPs, the HL group showed significantly higher air leakage compared to the TG group (p < 0.0001), and a higher PEEP resulted in greater air leakage (p < 0.0001). Air leakage with higher PEEP was greater in the HL group than in the TG group at ID 7.0 mm and 7.5 mm (p = 0.0467, p = 0.0045)
Conclusions
This study shows the superior sealing ability of the TG cuff during ventilation at various PEEPs.

Citations

Citations to this article as recorded by  
  • Impact of Low‐Volume, Low‐Pressure Tracheostomy Cuffs on Acute Mucosal Injury in Swine
    Alexandra J. Berges, Ioan A. Lina, Rafael Ospino, Hsiu‐Wen Tsai, Dacheng Ding, Jessica M. Izzi, Alexander T. Hillel
    Otolaryngology–Head and Neck Surgery.2022; 167(4): 716.     CrossRef
  • Tidal Volume Delivery and Endotracheal Tube Leak during Cardiopulmonary Resuscitation in Intubated Newborn Piglets with Hypoxic Cardiac Arrest Exposed to Different Modes of Ventilatory Support
    Marc R. Mendler, Claudia Weber, Mohammad A. Hassan, Li Huang, Benjamin Mayer, Helmut D. Hummler
    Neonatology.2017; 111(2): 100.     CrossRef

ACC : Acute and Critical Care